SwePub
Tyck till om SwePub Sök här!
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "LAR1:gu ;srt2:(2010);mspu:(article);pers:(Båth Magnus 1974);pers:(Jonasson Pernilla)"

Sökning: LAR1:gu > (2010) > Tidskriftsartikel > Båth Magnus 1974 > Jonasson Pernilla

  • Resultat 1-2 av 2
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Berner, Karin, et al. (författare)
  • Dose optimisation of double-contrast barium enema examinations.
  • 2010
  • Ingår i: Radiation protection dosimetry. - : Oxford University Press (OUP). - 1742-3406 .- 0144-8420. ; 139:1-3, s. 388-392
  • Tidskriftsartikel (refereegranskat)abstract
    • The purpose of the present work was to optimise the filtration and dose setting for double-contrast barium enema examinations using a Philips MultiDiagnost Eleva FD system. A phantom study was performed prior to a patient study. A CDRAD phantom was used in a study where copper and aluminium filtration, different detector doses and tube potentials were examined. The image quality was evaluated using the software CDRAD Analyser and the phantom dose was determined using the Monte Carlo-based software PCXMC. The original setting [100 % detector dose (660 nGy air kerma) and a total filtration of 3.5 mm Al, at 81 kVp] and two other settings identified by the phantom study (100 % detector dose and additional filtration of 1 mm Al and 0.2 mm Cu as well as 80 % detector dose and added filtration of 1 mm Al and 0.2 mm Cu) were included in the patient study. The patient study included 60 patients and up to 8 images from each patient. Six radiologists performed a visual grading characteristics study to evaluate the image quality. A four-step scale was used to judge the fulfillment of three image quality criteria. No overall statistical significant difference in image quality was found between the three settings (P > 0.05). The decrease in the effective dose for the settings in the patient study was 15 % when filtration was added and 34 % when both filtrations was added and detector dose was reduced. The study indicates that additional filtration of 1 mm Al and 0.2 mm Cu and a decrease in detector dose by 20 % from the original setting can be used in colon examinations with Philips MultiDiagnost Eleva FD to reduce the patient dose by 30 % without significantly affecting the image quality. For 20 exposures, this corresponds to a decrease in the effective dose from 1.6 to 1.1 mSv.
  •  
2.
  • Hansson, Jonny, et al. (författare)
  • A practical approach to prioritise among optimisation tasks in X-ray imaging: introducing the 4-bit concept
  • 2010
  • Ingår i: Radiation protection dosimetry. - : Oxford University Press (OUP). - 1742-3406 .- 0144-8420. ; 139:1-3, s. 393-399
  • Tidskriftsartikel (refereegranskat)abstract
    • According to European and national legislation, as well as international recommendations, X-ray examinations shall be optimised. However, with limited resources and hundreds of different types of X-ray examinations, it may be difficult to prioritise among the optimisation tasks at a radiology department. This work is focused on describing a method that can be used to determine the order of which the examinations should be optimised. In the Medical Exposure Directive from 1997, the European Commission prescribes the content of an optimisation process in relation to medical exposure. A reasonable interpretation of the directive is that the assurance of medical purpose for a justified examination is superior to the need of decreased radiation dose. This was used as a basis for developing a method for prioritisation among optimisation tasks. For each examination type, the following four yes/no questions are raised: (i) Is the present image quality unacceptable? (ii) Is the examination of particular importance? (iii) Is the radiation dose suspiciously high? (iv) Are there special dose level concerns, e.g. diagnostic reference levels? Arguing that a positive response to any of the four questions results in the examination being higher prioritised than otherwise and that the questions are labelled in order of decreasing relevance, it can be shown that the resulting flow chart, determining the order of which the examinations should be optimised, can be described by a 4-bit binary scale. In this way, each examination type is given a number from 0 to 15, a higher number corresponding to the examination being prioritised higher in the optimisation work. The method was applied to a general radiology department and resulted in a well-discriminated distribution of examinations prioritised for optimisation tasks. In conclusion, taking into account both medical outcome and potential risk, the proposed method can be used to determine the order in which examinations at a radiology department should be optimised.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-2 av 2
Typ av publikation
Typ av innehåll
refereegranskat (2)
Författare/redaktör
Månsson, Lars Gunnar ... (1)
Berner, Karin (1)
Cappelen-Smith, J (1)
Fogelstam, P (1)
visa fler...
Söderberg, Jonas (1)
Hansson, Jonny (1)
Sund, Patrik (1)
visa färre...
Lärosäte
Göteborgs universitet (2)
Språk
Engelska (2)
Forskningsämne (UKÄ/SCB)
Medicin och hälsovetenskap (2)
År

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy